Distributiestraat 73
4283 JN Giessen
The Netherlands

phone +31-(0)183-445050
mail info@mctec.nl

Bendtsen Tester – Roughness & Porosity

The Bendtsen Tester measures the air permeability and surface roughness according to Bendtsen. The sturdy and stable instrument has an industrial PC and a touchscreen. It is available with up to 3 heads: top roughness, bottom roughness and air permeability. The large display indicates single values and statistics. The measurement is performed quickly and automatically.

Productspecifications

Applicable standards:
ISO 8791-2
ISO 5636-1
ISO 5636-3

Features
• 1, 2, or 3 heads available in one device
• Standard measuring range: 25-5,000 ml/min (other measuring ranges also available, e.g. 1-5,000 ml/min)
• Simultaneous measurement with all 3 measuring heads
• Fully automated measuring cycle, sample is not influenced manually

Description
The sturdy and stable instrument possesses an industrial PC and a touchscreen. It is available with up to 3 heads: top roughness, bottom roughness and air permeability. The large display indicates single values and statistics. The measurement is performed quickly and automatically. The menu navigation
is clear.

Test description
Surface roughness:
The operator places the sample below the measuring head. By pushing the start button, the measuring head lowers onto the sample and gently clamps it. The test starts. As soon as the preset flow rate (e.g. 1.47 ml/min) is stable, the measuring value is displayed. The test time is adjustable from 1-40 seconds. After test performance, the sample is released. The measuring head returns to its initial position, and the device is ready for further measurements.

Air permeability:
The operator places the sample below the measuring head. By pushing the start button, the measuring head lowers onto the sample and gently clamps it. Air flows through the sample at a selected pressure (e.g. 1.47 kPa). As soon as the preset flow rate is stable, the measuring value is displayed in ml/min or μm/Pa*sec. The test time is adjustable from 1-40 seconds. Gurley values are calculated according to the standard and are displayed in Gurley seconds by default.

Specifications
• Measuring ranges:
– 25-5,000 ml flow rate
– Pressure difference: ± 0.5 % of the final value
• Measuring accuracy:
– Pressure differential: ± 0.005 kPa linear
– Flow: ± 0.5 % of the final value
– Accuracy of test duration: ± 0.5 s
• Flow sensors (optionally):
– 15-3,000 ml/min
– 5-1,000 ml/min
– 1-5,000 ml/min
• Air permeability:
– Measuring area: 10 ± 0.2 cm²
– Test duration: 1-40 seconds (adjustable)
– Display in ml/min or μm/Pa*sec
• Roughness:
– Measuring land: Ø 31.5 mm, 150 μm, 267 g
– Test duration: 1-40 seconds (adjustable)
• Touchscreen for easy handling
• Parallel printer interface
• Serial data interface (RS 232)
• Compatible to PaperLIMS and MiniStat
• Compatible to PTA-Line

Technical data
Electricity: 110–230 V, 50–60 Hz
Water supply: None
Compressed air: 400–600 kPa
Dimensions (WxDxH): 27x68x60 cm
Net weight:
– 1-head 23 kg
– 2-head 28 kg
– 3-head 34 kg
Gross weight:
– 1-head 38 kg
– 2-head 43 kg
– 3-head 49 kg

More information?